Chemicals / parameters Oxygen
Suitability Process liquids
Type Portable
System type Monitors
Features Alarm trip
Datalogging
Self cleaning
Weatherproof
 
 
 

The Q45D AutoClean dissolved oxygen monitor offers reliable dissolved oxygen measurement using a galvanic sensor with a rugged, 5 mil Teflon membrane to process and analyse the concentration of the oxygen.

The membrane resists mechanical abrasions and tears. It acts as a barrier to allow molecular oxygen to diffuse into the reaction cell where it is reduced, producing a current that is proportional to oxygen concentration.

The AutoClean system is available as an option to help reduce maintenance requirements. Supplying high pressure air blasts, it removes any build up of biological growth, foulants and contaminants from the sensor. The cleaning cycle frequency is adjustable to suit process requirements.

Technical information, electronic dissolved oxygen monitor:

  • Display range: 0 - 40.00 PPM
  • Accuracy: 0.2% of span or better
  • Display: Large 4 digit main display, 0.75" characters; 12 digit alpha-numeric second line display
  • Power: 16-35 VDC for loop-powered unit 115/230 VAC, 50/60 Hz., 10 VA max.; 9-Volt battery for battery operated portable

Technical information, dissolved oxygen sensor:

  • Sensor: Membrane-covered galvanic
  • Response time: 90% in 60 seconds (2 mil membrane); 90% in 180 seconds (5 mil membrane)
  • Temperature limits: 0-50° C.


The Q45D AutoClean dissolved oxygen monitor offers reliable dissolved oxygen measurement using a galvanic sensor with a rugged, 5 mil Teflon membrane to process and analyse the concentration of the oxygen.

The membrane resists mechanical abrasions and tears. It acts as a barrier to allow molecular oxygen to diffuse into the reaction cell where it is reduced, producing a current that is proportional to oxygen concentration.

The AutoClean system is available as an option to help reduce maintenance requirements. Supplying high pressure air blasts, it removes any build up of biological growth, foulants and contaminants from the sensor. The cleaning cycle frequency is adjustable to suit process requirements.

Technical information, electronic dissolved oxygen monitor:

  • Display range: 0 - 40.00 PPM
  • Accuracy: 0.2% of span or better
  • Display: Large 4 digit main display, 0.75" characters; 12 digit alpha-numeric second line display
  • Power: 16-35 VDC for loop-powered unit 115/230 VAC, 50/60 Hz., 10 VA max.; 9-Volt battery for battery operated portable

Technical information, dissolved oxygen sensor:

  • Sensor: Membrane-covered galvanic
  • Response time: 90% in 60 seconds (2 mil membrane); 90% in 180 seconds (5 mil membrane)
  • Temperature limits: 0-50° C.


Interested in this product?
Call 01457 832800  
OR
 
 
 
Downloads
  • Q45D dissolved oxygen monitor
    Q45D dissolved oxygen monitor
    6pp 632 KB
    Download
 
 
 
Enquire
 


Analytical Technology address and contact details

Analytical Technology, Inc.
Technology House
Unit 1 & 2, Gatehead Business Park
Delph New Road, Delph
Saddleworth
OL3 5DE
Tel: 01457 873318
View on map
 
 

Ask Analytical Technology about

Q45D AutoClean dissolved oxygen monitor   

Liquids measurement - products from other companies

Liquids measurement - all companies